1. BOD special socket-built-in power socket, convenient for the built-in instrument to power on.
2. Intelligent program controller-30-segment programming function to meet the needs of complex experiments.
3. Embedded printer-convenient for customers to print data.
4. Independent limit temperature alarm system-if the limit temperature is exceeded, the heating source is forced to stop, escorting your laboratory safety.
5. RS485 interface and special software-connect to a computer to export experimental data.
6. Test hole 25mm/50mm-can be used to test the actual temperature in the working room.
Technical Parameter
Optional Accessories:
Features
Project | 70F-I | 150F-I | 250F-I | 500F-I |
Voltage | AC220V 50HZ | |||
Temperature control range | 0~65℃ | |||
Constant temperature fluctuation | high temperature±0.3℃ Low temperature±0.5℃ | |||
Temperature uniformity | ±0.5℃ | |||
Temperature resolution | 0.1℃ | |||
compressor | Import Danfoss compressors from Denmark | |||
Freezer fan | Imported German ebian refrigerator fan | |||
Studio material | 304stainless steel | |||
refrigerating system | Fluorine-free environmental protection | |||
input power | 400W | 700W | 1150W | 2050W |
Liner size W×D×H(mm) | 450×320×500 | 480×390×780 | 580×500×850 | 700×700×1020 |
Dimensions W×D×H(mm) | 575×545×1070 | 605×625×1350 | 705×725×1525 | 825×995×1780 |
Nominal volume | 70L | 150L | 250L | 500L |
Carrying bracket (standard | 2pcs | 3pcs | ||
Timing range | 1~9999min |